Beckman Coulter AutoMate 2500 Series Instructions For Use Manual page 167

Sample processing system
Hide thumbs Also See for AutoMate 2500 Series:
Table of Contents

Advertisement

2
Wait 30 seconds after starting maintenance mode. The interlock releases the door.
Open the sliding door on the left of the recapper at the rear of the AutoMate.
3
Open the recapper hood.
4
Pull on the release knob of the guide table. Fold the guide table to the lower lock
position.
5
Press and hold Shift and advance the Parafilm gripper with
until the gripper reaches the front position and stops moving.
6
Inspect the Parafilm gripper for small pieces of Parafilm. Remove any Parafilm debris
you find.
7
Cut the film at the mark on the guide table and remove the film strip.
8
Press and hold Shift and move the Parafilm gripper back with
until the gripper reaches the end position and stops moving
9
Pull on the release knob of the guide table. Fold the guide table back to the upper locked
position.
10
Press
the brushes.
B67471B
to start manual film transport. Hold the button until you can see the film at
Recapper Operation and Maintenance
Recapper Troubleshooting