User Maintenance
Conveyor Belt Maintenance
Reference tube holder from the side and from below
3
Mark the reference tube holder position on the conveyor with a water resistant marker.
4
Each tube holder has a screw on the bottom. The screw has 0.5 mm gap to the white
chassis material. If this gap is clogged with residue, the system cannot identify the
conveyor belt positions.
5
If there is heavy residue buildup under the tube holders, remove all tube holders.
6
Use a vacuum cleaner to thoroughly clean the area.