Beckman Coulter AutoMate 2500 Series Instructions For Use Manual page 163

Sample processing system
Hide thumbs Also See for AutoMate 2500 Series:
Table of Contents

Advertisement

Figure 4.13
17
Pull on the release knob of the guide table, then fold the guide table back to the upper
lock position:
Figure 4.14
18
Press the
depressed until the film appears at the brushes (visual check).
19
Cut off the film with the
20
Pull on the release knob of the film cassette and push the roll to the lower lock position.
21
Close the recapper hood and the sliding door to the left of it.
22
On the touch screen, select OK to confirm that you have finished servicing the recapper.
23
Select Yes to confirm the initialization prompt that appears.
B67471B
button to initiate the manual film transport and hold the button
button and remove the strip from the brushes
Recapper Operation and Maintenance
Exchange the Film Cassette
4
4-11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Automate 1250Automate 2550Automate 1200

Table of Contents