Run the System
Aliquot Module Operation
Figure 2.14 Remove the wound-up roll
1
Bracket
4
Pinch roller
7
Printhead locking lever
Note
Dispose of the wound-up roll according to your laboratory procedure and all relevant
legal requirements. The roll can contain negative prints of patient data from the bar
code labels.
Note
The photo shows the full label roll loaded on the supply hub. When the labels are used
up, the empty liner is wound around the internal rewinder (3). Start here to remove the
roll of empty liner.
1
Turn the lever (7) counterclockwise to lift the printhead.
2
Cut the label strip and wind it fully around the internal rewinder (3).
3
Hold the internal rewinder (3) firmly and turn the knob (2) clockwise. The rewinder
spindle relaxes and the wound-up roll is released.
4
Remove the wound-up roll from the rewinder (3).
2-50
2
Turn knob
5
Rewind assist roller
3
Internal rewinder
6
Peeling edge
B67471B