Automatic Measurement In Harmonic Analysis Mode - AEMC OX III Series User Manual

100/200mhz portable oscilloscopes
Table of Contents

Advertisement

Applications (cont'd)
9. Automatic
measurement in
Harmonic Analysis
mode
Recall
Shop for AEMC products online at:
IX - 178
Initially, there must be one or two traces on channels CH1 to CH4.
- Only the channels (and not the functions) can be the subject of
harmonic analysis.
- In this mode, the time base is not adjustable.
Set the traces for channels CH1 to CH4 to the mode "Oscilloscope".
Using the key opposite, select the "Harmonic Analysis" mode.
The traces of channels CH1 and CH4 are shown in a dark color, while the
traces of CH2 and CH3 are light (or in the color of the trace).
Use the "Display" menu (see §. Display) to select the
harmonic breakdown required:
- the fundamental (F) and the first 15 harmonics
- the even harmonics (2 to 30)
- the odd harmonics (3 to 31)
The selection chosen appears under the breakdown display.
The "SIGNAL" table under the breakdown provides information on:
- the active channel(s) (CH1 to CH4)
- the RMS voltage of the signal in V
- harmonic distortion rate (in %) of the signal
The " " symbol, present on the fundamental (F) or one of the
harmonics and modifiable using the stylus, means that measurements
can be made on the harmonic selected.