Oscilloscope Mode (cont'd)
Pulse
Example
Shop for AEMC products online at:
IV - 70
Trigger selection on pulse width.
In all cases the effective triggering occurs on the end of pulse edge.
t > T1
t < T2
t > T1 and t < T2
t < T1 or t > T2
The choice of the
analysis:
• edge
defines a pulse between
• edge
defines a pulse between
Signal injected on CH1: 1 pulse of 50 µs (repetitive or not)
The oscilloscope triggers when it sees a pulse whose duration is greater
than specified value (25.0 μs + tolerance). The trigger occurs on the trailing
edge of the pulse.
The trigger occurs on the ascending edge of the pulse, but is effective on the
trailing edge.
ShopAEMC
www.
Oscilloscope Mode - The "Trigger" Menu
triggers if pulse duration is greater than specified value T1
triggers if pulse duration is less than specified value T1
triggers if pulse duration is within interval specified by
values T1 and T2
triggers if pulse duration is outside interval specified by
values T1 and T2
or
edge in the "Main" tab defines the limits of the
.com
and
and
1.877.766.5412
SCOPIX