Multimeter Mode (cont'd)
The "Trig" Menu
Source/Level
Source
Type
Level 1
Level 2
min duration >
Shop for AEMC products online at:
V - 110
Multimeter Mode - The "Trigger", "Horizontal" and "Display" Menus
Selection of trigger type and level on each channel. Triggering takes place if a
condition described by a line of the "Trigger" table is verified.
The trigger level should be defined in the channel measurement dynamic.
Trigger leads to the logging of the instant and trigger characteristics.
The events logged can be accessed from the "Disp."
The channel number.
The type of trigger for each channel.
There are several possible types:
None (no triggering)
Lower than
Higher than
Lower/Higher
Exterior
In "Multimeter" mode, several conditions are monitored simultaneously.
Adjustment of the main trigger level using the stylus.
Adjustment of the auxiliary trigger level using the stylus.
This tab is enabled only if the "External" trigger type is selected.
The fault will be evidenced if the fault condition defined by the type and
levels is present for a parameterisable period from 480 ms to 670 ks
according to the selected recording duration (48 ms to 670 ks if your
instrument is equipped wth the EXTENDED ACQUISITION MEMORY
option).
ShopAEMC
www.
"Defaults" menu.
.com
1.877.766.5412
SCOPIX