AEMC OX III Series User Manual page 172

100/200mhz portable oscilloscopes
Table of Contents

Advertisement

Applications (cont'd)
HX0030 probe
compensation
Shop for AEMC products online at:
IX - 172
If necessary:
• Modify the trigger level using the stylus:
- by moving the symbol T (Trigger) on the screen. The value of the trigger
level is indicated in the bottom right-hand corner of the screen.
- or through the trigger parameters menu:
Trig. Menu
• Modify the vertical positioning of the curve:
- by using the stylus to move symbol 1, on the left of the screen.
- or by using the keys opposite.
The key opposite is used to make these adjustments automatically.
Adjust the low-frequency compensation of the probe so that the signal plateau is
horizontal (see figure below).
Over-compensated probe
Compensated probe
Under-compensated probe
Adjust the screw on the Probix HX0030 probe to adjust the compensation.