AEMC OX III Series User Manual page 109

100/200mhz portable oscilloscopes
Table of Contents

Advertisement

Multimeter Mode - The "Vertical" Menu
Multimeter Mode (cont'd)
Autorange
When this option is selected, the measurement range changes
automatically.

The "" symbol shows that it is active.
If the option is disabled, the range can be modified manually, using the
keys opposite or the "Range" menu, depending on the type of
measurement being performed.
Coupling
Modification of the AC, DC or AC+DC coupling for amplitude measurement.
AC: AC voltage measurement
DC: DC voltage measurement
AC + DC: Measurement of AC voltage with a DC component
For AC and AC+DC measurements, the menu "Display
be used to display the frequency of the signal in the secondary display field.
The "
The coupling is updated in the modified channel parameter display zone.
Bandwidth limit
If the channel measures an AC or AC+DC voltage (see Coupling), it is
possible to filter the signal with a low-pass analogical filter whose cut-off
frequency is 5 kHz.
The other proposed filter is a digital filter to 625Hz, if this filter is selected,
the analog filter at 5kHz is also activated.
The digital filter characteristics are:
Low-pass filter
Cutoff frequency ..................................................................... 625Hz
Order ............................................................................................. 94
Passband ripple ....................................................................... 0.5dB
Transition band ........................................................................... 0.02
Stopband attenuation ............................................................ 50.0dB
SCOPIX
Shop for AEMC products online at:
" symbol indicates the type of coupling selected.
ShopAEMC
www.
Frequency" can
.com
1.877.766.5412
V - 109

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ox 7202 iiiOx 7102 iiiOx 7104 iiiOx 7204 iii

Table of Contents