Technical Specifications
Mains « Harmonics » Analysis Mode
Display of "Harmonics"
Even harmonics
Odd harmonics
All harmonics
Fundamental Frequency of the
signal analyzed
Measurement accuracy
Level of Fundamental
Level of Harmonics
Harmonic Distortion
« Recorder » Mode
Recording period
Sampling rate
Fault capture
Triggering
Trigger event period
Min. value of the period
Max. value of the period
Display
Vertical, horizontal accuracy
Shop for AEMC products online at:
X - 214
Selection of the page in the "Display" menu
2 to 60 + Fundamental on 2 pages
3 to 31 + Fundamental on 2 pages
2 to 61 + Fundamental on 4 pages
40 to 450 Hz
± 2 % ± +10 of Reading
± 3 % ± +10 of Reading
± 4 %
Oscilloscope not equipped
with the
EXTENDED ACQUISITION
MEMORY option
from 800 µs to 17 mins 51 secs
10 faults in memory up to
500 faults in files
Triggering search by sample analysis;
Trigger if Detection of:
•
Signal above threshold
•
Signal below threshold
•
Signal below or above threshold
•
Signal outside of two defined limits
Trigger if the condition is verified during a parameterizable period:
Recording period / 625
Recording period / 4
Search for minimum and maximum
Identical specifications to those in "Oscilloscope" mode
ShopAEMC
www.
Oscilloscope equipped with the
EXTENDED ACQUISITION
from 2 seconds to 1 month
from 40 µs to 53.5 secs
100 faults in memory up to
Recording period / 12500
Fault search
.com
1.877.766.5412
Technical Specifications
MEMORY option
200 faults in files
Recording period / 4
SCOPIX