AEMC OX III Series User Manual page 201

100/200mhz portable oscilloscopes
Table of Contents

Advertisement

Technical Specifications
Technical Specifications (cont'd)
« Oscilloscope » Function
Horizontal deflection (time base)
Characteristics
Time base ranges
Time base accuracy
Sampling rate
Time measurement accuracy
Horizontal ZOOM
Mode XY
Phase error
Representation
Fast Fourier Transform
Shop for AEMC products online at:
SCOPIX
III
OX 7102
- OX 7104
without the
EXTENDED ACQUISITION MEMORY
option
35 ranges, from 1 ns to 200 s/div.
± [50 ppm + max (500 ps, 1 sample)]
2.5 GS/sec. in real time
50 GS/sec. with repetitive signal
± [(0.02 div.) x (time/div.) + 0,01 x reading + 1 ns]
Zoom factor : from x 1 to x 5
recording memory capacity :
2,500 samples per channel
In ZOOM and normal modes : the same sequence of time base range is used.
The horizontal resolution of the screen is 500 samples for 10 divisions.
The bandwidths are identical in X and in Y. (See §. Vertical deflection).
calculation on the traces present in the screen area
dynamic refreshment as a function of the signal observed in RUN mode
windowing: rectangle, Hamming, Hanning, Blackman
scales: logarithmic or linear
automatic adjustment with autoset function
ShopAEMC
www.
III
- OX 7202
EXTENDED ACQUISITION MEMORY
2.5 GS/sec. in real time
100 GS/sec. with repetitive signal
Zoom factor : from x 1 to x 100
recording memory capacity :
50,000 samples per channel
< 3°
temporal or frequential (FFT)
.com
1.877.766.5412
III
III
- OX 7204
with the
option
X - 201

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ox 7202 iiiOx 7102 iiiOx 7104 iiiOx 7204 iii

Table of Contents