Automatic Measurement In Harmonic Analysis Mode - AEMC OX7102-CII User Manual

Portable oscilloscope ox-cii series
Table of Contents

Advertisement

Applications (cont'd)
9. Automatic
measurement in
Harmonic Analysis
mode
Recall
IX - 164
Initially, there must be one or two traces on channels CH1 to CH4.
- Only the channels (and not the functions) can be the subject of
harmonic analysis.
- In this mode, the time base is not adjustable.
Set the traces for channels CH1 to CH4 to the mode "Oscilloscope".
Using the key opposite, select the "Harmonic Analysis" mode.
Th
e traces of
channels CH1 and CH4 are s
trac
es of CH2
and CH3 are light (or in the c
Use the "Display" menu (see §. Display) to select the
harm
onic b
reakdown required:
- the fundamental (F) and the first 15 harmonics
-
the even harmonics (2 to 30)
- the odd harmonics (3 to 31)
The selection chosen appears under the breakdown display.
The "SIGNAL" table under the breakdown provides information on:
- the active channel(s) (CH1 to CH4)
- the RMS voltage of the signal in V
- harmonic distortion rate (in %) of the signal
The " " symbol, present on the fundamental (F) or one of the
harmonics and modifiable using the stylus, means that measurements
can be made on the harmonic selected.
hown in a dark co
lour, while the
olour of the trace).