Table 3-1
IEEE 488.2 Common Commands (continued)
Mnemonic
*OPC?
*PRE <NRf>
*PRE?
*RCL <NRf>
*RST
*SAV <NRf>
*SRE <NRf>
*SRE?
*STB?
*TRG
*TST?
*WAI
Note: Pending operations include measurements in progress.
Keysight 53147A/148A/149A Operating and Programming Guide
Command Name
Operation Complete Query
Parallel Poll Enable Register Sets the value of the Parallel Poll Enable register.
Parallel Poll Enable Register
Query
Recall
Reset
Save
Service Request Enable
Service Request Enable
Query
Status Byte Query
Trigger
Self-Test Query
Wait-to-Continue
Function
Places an ASCII "1" in the Output Queue when all pending
operations (see Note at end of table) are completed.
Queries the value of the Parallel Poll Enable register.
Restores the state of the instrument's user settings from a copy
stored in local non-volatile memory
(0 through 9 are valid memory registers).
Resets the instrument to a known state, as defined in this manual.
Stores the current state of the instrument's user settings in local
non-volatile memory (0 through 9 are valid memory registers).
Sets the Service Request Enable register.
Queries the Service Request Enable register.
Queries the Status Byte and Master Summary Status bit.
This trigger command is the device-specific analog of the IEEE
488.1 defined GET. It initiates the action specified by the *DDT
command.
Executes an internal self-test and reports the results.
Makes the instrument wait until all pending operations (see Note)
are completed before executing commands that follow the *WAI
command.