:TRIGger Subsystem

:TRIGger[SEQuence | STARt]:HOLDoff . . . <numeric_value>
COMMAND
When INIT:CONT is ON, this command determines the rate at which
measurements are made by setting a delay between measurements. Its setting
corresponds to the front-panel rate setting as follows:
0.0 = FAST; 0.5 = MEDium; 1.0 = SLOW
<NUMERIC_
VALUE> RATE
– *RST: 0.0
COMMENTS
– The only settings that are accepted are the three listed above.
– The measurement rate is also affected by the Resolution setting.
– When INIT:CONT is OFF, the holdoff has no effect.
– There is no setting for this command that corresponds to the front-panel rate
Shift + Rate
RELATED
FRONT-PANEL
KEYS
Keysight 53147A/148A/149A Operating and Programming Guide
setting of HOLD. The HOLD mode is entered through *RST or INIT:CONT OFF.
The instrument then stops making measurements until it is triggered or until it
receives INIT:CONT ON.