Status:questionable Subtree; Figure 3-19 The Questionable Data/Signal Status Register - Keysight 53147A Operating And Programming Manual

Microwave frequency counter/power meter/dvm
Table of Contents

Advertisement

:STATus:QUEStionable Subtree

The :STATus:QUEStionable subtree commands allow you to examine the status of
the instrument monitored by the Questionable Data/Signal Status Register Group,
shown in
register, two transition registers, an event register, and an enable register. The
commands in this subtree allow you to control and monitor these registers.
See the sections titled
Status Register Group"
Questionable Data/Signal Status Register Group.
Hardware Summary
Command Warning
Figure 3-19
:STATus:QUEStionable:CONDition?
COMMAND
Queries the status of the Questionable Data Condition Status Register.
– Numeric data transferred as ASCII bytes in <NR1> format.
QUERY
RESPONSE
– Range is 0 to 32,767.
– The query response value is an integer formed by the binary weighting of the
– The Questionable Data Condition Status Register is cleared at
COMMENTS
– Bits are not cleared when read.
None
RELATED
FRONT-PANEL
KEYS
Keysight 53147A/148A/149A Operating and Programming Guide
Figure
3-19. The Questionable Status Group consists of a condition
"Operation Status Register Group"
Not Used
( 8 )
Power
Not Used
( 32 )
Frequency
Not Used
(4096)
Not Used
( 16384 )
Not Used
The Questionable Data/Signal Status Register Group
bits. The value of unused bits is zero.
power-on.
on pages 184 and 186 for a detailed description of the
Bits
0 to 2
3
4
5
6 to 11
12
13
14
15
Programming
and
"Questionable Data