Maintenance
Cleaning the Electrodes, Pre-punchers and Insulation Plate

Reinstalling the Electrodes and the Insulation Plate

1 Reinstall one electrode in the outlet electrode position on the insulation
2 Thread the electrode into the outlet position on the insulation plate by hand
3 Tighten the electrode with the 12 mm hexagonal socket screwdriver.
4 Holding the insulation plate by the cassette holder, slide the insulation
3. Left shift
2. Rotation
1. Insert
Figure 67
210
plate.
first.
Do not overtighten. Do not bend the electrodes.
plate under the detector (1.), then tilt it up into place (2.) and push it
completely to the left (3.).
Installation of the insulation plate
Agilent 7100 Capillary Electrophoresis System User Manual