Cassette front
Cassette back
Figure 32
Installing the Capillary
1 Hold the capillary at the alignment interface.
2 Insert the alignment interface into the alignment interface holder on the
3 Make sure the capillary slides into the capillary guiding slit and a guiding
4 Wind the capillary around the reel in the center of the capillary cassette.
Agilent 7100 Capillary Electrophoresis System User Manual
Opening the cassette lock
cassette. Make sure the flat side of the alignment interface is at the outer
edge of the cassette, see
hook above the area of the alignment interface.
The pins surrounding the reel have grooves to secure the capillary.
Using the Agilent 7100 Capillary Electrophoresis System
Figure 33
on page 130.
Capillary Electrophoresis, CE
5
129