Reinstall the Capillary Cassette

See

Reinstall the Front Cover

1 Make sure the front cover screws have a horizontal orientation. They can
2 Keep the front door of the cabinet slightly open.
3 Start mounting the front cover by inserting it at the lower end first.
4 Carefully push the front cover into the main frame.
5 Close the lid of the instrument.
Agilent 7100 Capillary Electrophoresis System User Manual
Cleaning the Electrodes, Pre-punchers and Insulation Plate
"Inserting the Cassette"
only click into place in this orientation.
You might need to pay special care on the right side due to a connector.
on page 132.
Maintenance
8
215