Maintenance
Cleaning the Electrodes, Pre-punchers and Insulation Plate

Remove the Covers

1 Open the top cover.
2 Remove the capillary cassette (see
3 Turn the front cover screws by a quarter of a turn to vertical orientation.
4 Open the door to the cabinet a little.
5 Carefully pull the front cover away from the main frame.
Screws
Front cover
Connector
Figure 57
198
the Agilent CE Instrument"
They are not meant to be removed completely.
Pull steadily on both sides. You will need slightly more force to remove the
right side due to a connector. Turn the screws back to horizontal
orientation.
Removing the front cover
"Removing the Capillary Cassette from
on page 121)
Agilent 7100 Capillary Electrophoresis System User Manual