Using an Injection Program
The injection table is used for advanced injection tasks such as:
• Injection from different vials, for example ITP.
• Multiple injection modes, for example spiking.
• Injecting a buffer plug after the sample to prevent sample loss after
The online help provides more information on this.
Timetable
In the CE Timetable, present on the right-hand side of the CE Method Setup dialog
you can choose to set and change values for the following parameters at a
certain time during the run:
• Voltage
• Current
• Power
• Pressure
• Inlet vial
• Outlet vial
• Cassette temperature
• Lower alarm limit for current
• High pressure (only in CEC mode)
Time programming of these parameters can be useful for example to:
• Define gradients (for voltage or temperature).
• Change the buffer during the analysis (inlet and outlet).
• Apply pressure during the analysis.
• Change the outlet vial during analysis.
• Change the alarm limit for current for example, when changing the buffer
Agilent 7100 Capillary Electrophoresis System User Manual
applying voltage.
system during the analysis.
Using the Agilent 7100 Capillary Electrophoresis System
Capillary Electrophoresis, CE
5
105