Cleaning the Electrodes, Pre-punchers and Insulation Plate
4 Carefully lift the red high voltage wire. The inlet electrode will come out.
5 Slowly slide the insulation plate first to the right (1.), then tilt it up (2. and
1. Right shift
2. Rotation
3. Take out
Figure 60
It is not possible to remove the outlet electrode without taking out the insulation plate.
N O T E
6 Remove the outlet electrode from the insulation plate using the 12 mm
200
3.) and finally take it out.
Make sure that the outlet electrode is not touching the foam of the tray
cooling or the lift station.
Removing the insulation plate
hexagonal socket screwdriver.
Agilent 7100 Capillary Electrophoresis System User Manual