Timetable controlled changes in the pressure settings are:
• OFF: Pressure usage is switched off.
• Internal, pressure (-100 mbar to +100 mbar): Can be applied to the inlet
• Flush: Standard flush procedure is applied with an internally generated
• External: Allows pressurizing the inlet vial and/or the outlet vial. External
• Raw External Pressure: Allows pressurizing the inlet vial, and/or the outlet

CE+p Control Using the CE Diagram

High pressure may be applied independently of the method for capillary
conditioning or cleaning by clicking the external pressure icon in the CE
diagram (see
to appear (see
pressure to the inlet, outlet or both vials, and also whether to simultaneously
apply a voltage of -30 to 30 kV for a given time.
Agilent 7100 Capillary Electrophoresis System User Manual
vial.
pressure of approximately 950 mbar.
pressure is reduced to the set value between 2 and 12 bar and applied in a
controlled manner.
vial. The external pressure up to 15 bar can be applied directly without
reduction.
Figure 39
Figure 41
Using the Agilent 7100 Capillary Electrophoresis System
Capillary Electrophoresis, Plus High Pressure CE+p
on page 142). This will cause a parameter setting field
on page 145). From this field you may chose to apply
5
147