5
Using the Agilent 7100 Capillary Electrophoresis System
Capillary Electrophoresis, Plus High Pressure CE+p

Injection Options with CE+p

The installed high pressure supply may also be used for injection purposes
using the Timetable option in the Injection field of a method setup.
From this menu you may select the High pressure flush command menu during
injection with the same parameter choices that are available during
Preconditioning.
This may be used to perform a pressure injection while using a capillary filled
with replaceable gel.
Please note that the external pressure is not as precise as the injection pressure of up to
N O T E
100 mbar!

Timetable Options with CE+p

The installed high pressure supply may also be used in the general timetable of
a method setup. The Pressure command is extended by the External pressure
functionality, see
Figure 42
146
Figure 42
on page 146.
Pressure options in CE+p mode