Agilent Technologies 7100 User Manual page 74

Capillary electrophoresis system
Hide thumbs Also See for 7100:

Advertisement

4
LAN Configuration
Manual Configuration
5 Use "/" and press [Enter] to list the current settings.
Figure 11
6 Change the IP address (in this example 134.40.27.99) and type "/" to list the
If the initialization mode switch is changed now to "Using Stored" mode, the instrument
N O T E
will take over the stored settings when the instrument is re-booted. In the example above, it
would be 134.40.27.99.
74
Telnet - current settings in "Using Stored" mode
current settings.
Information about the LAN interface
MAC address, initialization mode
Initialization mode is Using Stored
active TCP/IP settings
TCP/IP status - here ready
connected to a PC with controller software
(for example Agilent ChemStation), here not
connected
Agilent 7100 Capillary Electrophoresis System User Manual