TCP/IP Parameter Configuration

To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. These parameters are:
• IP address
• Subnet mask
• Default gateway
The TCP/IP parameters can be configured by the following methods:
• by automatically requesting the parameters from a network-based BOOTP
• by manually setting the parameters using Telnet
The LAN interface differentiates between several initialization modes. The
initialization mode (abbreviation: 'init mode') defines how to determine the
active TCP/IP parameters after power-on. The parameters may be derived
from a Bootp cycle, non-volatile memory or initialized with known default
values. The initialization mode is selected by the configuration switch, see
Table 17
Agilent 7100 Capillary Electrophoresis System User Manual
service (using the so-called bootstrap protocol), see
Service"
on page 75
on page 64.
LAN Configuration

TCP/IP Parameter Configuration