LAN Configuration
Initialization Mode Selection
Table 19
IP address:
Subnet mask:
Default gateway
Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the CE instrument must be in the same
subnet.
The user may open a Telnet session using the default IP address and change
the parameters stored in the non-volatile memory of the CE instrument. He
may then close the session, select the initialization mode "Using Stored",
power-on again and establish the TCP/IP connection using the new
parameters.
When the 7100 CE instrument is wired to the PC directly (for example using a
cross-over cable or a local hub), that is separate from the local area network,
the user may simply keep the default parameters to establish the TCP/IP
connection.
In the "Using Default" mode, the parameters stored in the memory of the 7100 Capillary
N O T E
Electrophoresis instrument are not cleared automatically. If not changed by the user, they
are still available when you switch back to the mode "Using Stored".
68
Using Default parameters
192.168.254.11
255.255.255.0
Not specified
Agilent 7100 Capillary Electrophoresis System User Manual