Agilent 7890A GC
Safety Manual
Agilent Technologies

Advertisement

loading

Summary of Contents for Agilent Technologies 7890A GC

  • Page 1 Agilent 7890A GC Safety Manual Agilent Technologies...
  • Page 2 Notices © Agilent Technologies, Inc. 2007 Warranty Safety Notices No part of this manual may be reproduced in The material contained in this docu- any form or by any means (including elec- ment is provided “as is,” and is sub-...
  • Page 3: Table Of Contents

    Agilent 7890A GC Safety Manual Introduction Important Safety Warnings 4 Hydrogen Safety 7 Micro-Electron Capture Detector (µECD) 14 Fuses and batteries 17 Safety and Regulatory Certifications 18 Cleaning 22 Recycling the Product 22 Agilent Technologies...
  • Page 4: Important Safety Warnings

    Important Safety Warnings Before moving on, there are several important safety notices that you should always keep in mind when using the Agilent 7890A GC. When handling/using chemicals for preparation or use within the GC, all applicable WA RN ING local and national laboratory safety practices must be followed.
  • Page 5 Whenever possible, cool the part of the instrument that you will be maintaining before you begin working on it. Be careful when working behind the instrument. During cool-down cycles, the GC WA RN ING emits hot exhaust which can cause burns. 7890A GC Safety Manual...
  • Page 6 Objects passing through the oven door seal can cause thermal leaks which create WA RN ING hazardous hot spots which cause burns and melt equipment. Do not allow wiring or temperature probes to pass through the oven door jam. Agilent recommends using one of the access holes. 7890A GC Safety Manual...
  • Page 7: Hydrogen Safety

    GC assumes that a leak has occurred and declares a hydrogen safety shutdown. The effects are: 7890A GC Safety Manual...
  • Page 8 The H streams must be configured for hydrogen so that the GC is aware of hydrogen use. 7890A GC Safety Manual...
  • Page 9 It can also be shut down accidentally by an internal or external failure. A mass spectrometer shutdown does not shut off the flow of carrier gas. As a result, hydrogen may slowly accumulate in the mass spectrometer. 7890A GC Safety Manual...
  • Page 10 Different GCs react in different ways. If a 7890A GC equipped with Electronic Pressure Control (EPC) is shut off, the EPC stops the flow of carrier gas. If the carrier flow is not under EPC control, the flow increases to its maximum.
  • Page 11 You MUST make sure the front side-plate thumbscrew is fastened finger-tight. Do not overtighten the thumbscrew; it can cause air leaks. Failure to secure your MSD as described above greatly increases the chance of WA RN ING personal injury in the event of an explosion. 7890A GC Safety Manual...
  • Page 12 Environmental Health and Safety (EHS) requirements. Always check for leaks after changing a tank or servicing the gas lines. Always make sure the vent line is vented into a fume hood. 7890A GC Safety Manual...
  • Page 13 When measuring gas flows on a detector using hydrogen for the detector flame or carrier gas, measure the hydrogen flow separately. Never allow an air stream to enter when hydrogen is present in the flow meter. 7890A GC Safety Manual...
  • Page 14: Micro-Electron Capture Detector (Μecd)

    These materials include oxidizing compounds, acids, wet halogens, wet nitric acid, ammonium hydroxide, hydrogen sulfide, PCPs, and carbon monoxide. This list is not exhaustive but indicates the kinds of compounds that may cause damage to Ni detectors. 7890A GC Safety Manual...
  • Page 15 Wear disposable plastic gloves and observe normal laboratory safety precautions. Contact your local Agilent Technologies sales office or distributor for ECD disposal instructions Include a letter stating the condition of abuse.
  • Page 16 For other countries, consult with the appropriate agency for equivalent requirements. Agilent Technologies recommends a vent line internal diameter of 6 mm (1/4-inch) or greater. With a line of this diameter, the length is not critical. 7890A GC Safety Manual...
  • Page 17: Fuses And Batteries

    20 A, 250 Vac, IEC 127 type f (non-time delay), ceramic body F3, F4 8 A, 250 Vac, IEC type f (non-time delay), glass body Table 3 Logic board battery Battery designation Battery rating and type 3-volt lithium-poly carbon battery, .048A-HR, Panasonic model BR 1225 7890A GC Safety Manual...
  • Page 18: Safety And Regulatory Certifications

    This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB—001 du Canada. The Agilent 7890A GC is designed and manufactured under a quality system registered to ISO 9001. Instructions for Disposal of Waste Equipment by Users in the European Union.
  • Page 19 Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
  • Page 20 4 Make sure that all peripheral devices are also certified. 5 Make sure that appropriate cables are used to connect the device to peripheral equipment. 6 Consult your equipment dealer, Agilent Technologies, or an experienced technician for assistance. 7 Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to operate the equipment.
  • Page 21 Sound Emission Certification for Federal Republic of Germany Sound pressure Sound pressure Lp < 70 dB(A) according to DIN-EN 27779. Schalldruckpegel Schalldruckpegel LP < 70 dB(A) nach DIN-EN 27779. 7890A GC Safety Manual...
  • Page 22: Cleaning

    Agilent Technologies Cleaning To clean the unit, disconnect the power and wipe down with a damp, lint-free cloth. Recycling the Product For recycling, contact your local Agilent sales office. © Agilent Technologies, Inc. Printed in USA, February 2007...