LAN Configuration

Storing the Settings Permanently with Bootp

Storing the Settings Permanently with Bootp

If you want to change parameters of the CE instrument using Bootp follow the
instructions below.
1 Turn off the CE instrument.
2 Change the CE instrument's settings for the configuration switch to "Bootp
3 Start the Agilent Bootp Service (see
4 If required, modify the parameters for the CE instrument according to your
5 Press OK to exit the Bootp manager.
6 Now turn on the CE instrument and view the Bootp server window. After
7 Close the Agilent Bootp Service and turn off the CE instrument.
8 Change the settings of the CE instrument's configuration switch to "Using
9 Power cycle the CE instrument. The CE instrument can be accessed now via
70
& Store" mode, see
Table 18
and open its window.
needs using the existing configuration.
some time the Agilent Bootp Service will display the request from the LAN
interface. The parameters are now stored permanently in the non-volatile
memory of the CE instrument.
Stored" mode, see
Table 18
LAN without the Agilent Bootp Service, refer to
page 75.
on page 65.
"Agilent Bootp Service"
on page 65.
Agilent 7100 Capillary Electrophoresis System User Manual
on page 75)
"Agilent Bootp Service"
on