Timetable Options with CEC

The installed high pressure supply may also be used in the general timetable of
a method setup. The Pressure command is extended by the External pressure
functionality, see
Figure 49
Timetable-controlled changes in the pressure settings are:
• OFF: Pressure usage is switched off.
• Internal, pressure (-100 mbar to +100 mbar): Can be applied to the inlet
• Flush: Standard flush procedure is applied with an internally generated
• External: Allows pressurizing the inlet vial and/or the outlet vial. External
Agilent 7100 Capillary Electrophoresis System User Manual
Figure 49
Pressure options in CEC mode
vial.
pressure of approximately 950 mbar.
pressure is reduced to the set value between 2 and 12 bar and applied in a
controlled manner.
Using the Agilent 7100 Capillary Electrophoresis System
Capillary Electrochromatography, CEC
on page 157.