10 Tighten the slotted screws in the front of the liquid handling module.
3. Front screws of the
liquid handling module
2. Put in bottles
1. Push liquid handling
module in
Figure 70
11 Reconnect the line power, drainage, external pressure supply, the
12 Reestablish the connection to the instrument from the Lab Advisor
No high voltage can be applied, if the front cover is removed. Touching the electrodes inside
N O T E
the instrument is safe. However, under the following circumstances (usage of the
alignment center), slow lifter movements with a low force are possible with the following
alignment procedure. When you start any lifter movement your hands should be outside the
instrument.
Agilent 7100 Capillary Electrophoresis System User Manual
Cleaning the Electrodes, Pre-punchers and Insulation Plate
Liquid handling module
connectors for an external cooling bath (if applicable), and switch on the
instrument.
software. Select Start monitoring.
Maintenance
8
213