Removing the Capillary Cassette from the Agilent CE Instrument
Your Agilent CE instrument comes with an empty capillary cassette installed.
To remove the capillary cassette from the Agilent CE instrument, you have to
do the following steps:
1 Choose the cassette icon in the CE diagram and select Change Cassette...
2 Open the top cover to access the cassette. Press a spring located in the
If you do not lower the vials from the loaded position, buffer, which may be transferred via
N O T E
the capillary into the electrode inserts, precipitates and may block the electrodes or
damage the capillary.
3 Move the cassette guide to the front.
Agilent 7100 Capillary Electrophoresis System User Manual
from the menu. This removes the buffer vials from the ends of the capillary
abd places them back to their position in the tray. This will take
approximately 5 seconds.
middle of the front lid to release the latch.
cassette," on page 122 shows the position of an installed cassette.
Using the Agilent 7100 Capillary Electrophoresis System
Capillary Electrophoresis, CE
Figure
26, "Position of installed