Accessing the Electrodes

1 Use a screwdriver to unscrew the two metal screws holding the insulation
2 Use a screwdriver to remove the screw that secures the blue ground cable.
Metal screws
Insulation plate
Figure 58
3 Use the 12 mm hexagonal socket screwdriver to unscrew the inlet electrode.
12 mm hexagonal
socket screwdriver
High voltage cable, red
Ground cable screw
Ground cable, blue
Figure 59
Agilent 7100 Capillary Electrophoresis System User Manual
Cleaning the Electrodes, Pre-punchers and Insulation Plate
plate.
Insulation plate
Removing the ground cable and the inlet electrode cable
Maintenance
8
199