Cleaning the Electrodes, Pre-punchers and Insulation Plate

Cleaning the Pre-punchers

• Check for damaged or bent tips before cleaning the pre-punchers.
• Check for grooves in the funnels of the pre-punchers.
Risk of injury due to sharp tips
WA R N I N G
The pre-punchers have very sharp tips.
• Be careful when touching the tips of the pre-punchers.
1 Flush the pre-punchers with water. Watch for salt crystals. All salt crystals
Figure 65
2 Flush the pre-punchers with water and isopropanol.
3 Use compressed air to dry the pre-punchers and remove all traces of liquid.
206
If the tips are damaged, replace the pre-punchers.
Grooves are a sign of improper alignment of the electrodes and may lead to
improper sealing. Improper sealing may lead to problems during flushing,
injection or CEC applications. Use the alignment center of the Lab Advisor
to align and check the installation. If the funnel shows grooves, replace the
pre-punchers.
must be removed. Take particular care with the teflon tube connector.
Flushing pre-punchers
Do not forget to dry the connected tubes as well.