Hardware Information
7100 Capillary Electrophoresis Instrument Electronics
All electrical connectors are placed on the right side of the instrument, see
Figure 53
Available Connectors:
• Power connector, Fuse netfilter: 2 fuses, 250 VAC, T6.3AH, (part number
• LAN connector (7100 to controlling PC)
• CAN connectors (7100 to additional Agilent 1200 Series Module)
• USB connector (for future use)
• RS232 connector (serial interface to a computer. This connection has a
• APG remote connector (Agilent Technologies Analytical Products Group
• Analog out (BNC connector, 0 to 0.1 V or 0 to 1 V)
• Analog in (BNC connector, 0 to 1 V, shield is analog - and center is analog +)
• 8-bit configuration switch (see
Fuses
• Fuse 1 (Peltier) T6.3AL, 250 VAC, (part number 2110-0623)
• Fuse 2 (HV supply, CE detector board) T3.15 AL, 250 VAC, (part number
• Fuse 3 (CE sampler board with valve and lifter control) T3.15 AL, 250 VAC,
• Fuse 4 (heater, optional), T3.15 AL, 250 VAC, (part number 2110-1417)
• Fuse 5 (processor, +5 V, +15 V, -15 V supplies on mainboard, air pump, fans,
• Fuse 6 (DAD) T3.15 AL, 250 VAC, (part number 2110-1417)
174
on page 173.
2110-1018)
special out pin and is not compatible for connecting printers or plotters. It
is called "null modem cable" with full handshaking and the wiring is
established between pins 1-1, 2-3, 3-2,4-6, 5-5, 6-4, 7-8, 8-7 and 9-9.)
remote connector to be connected to Agilent modules, for example to MS
system)
page 61
2110-1417)
(part number 2110-1417)
CE temperature sensor board and cassette fan) T3.15 AL, 250 VAC, (part
number 2110-1417)