Troubleshooting and Diagnostics
6
CE Problem Solving
Leaks
The leak sensor in the Agilent CE instrument is located in the lower right part
of the instrument, see
Figure
50.
Leak sensor
Figure 50
Leak sensor position
When the leak sensor is activated:
• The error LED on the front of the instrument is illuminated.
• The CE status displays an error.
• A logbook entry is created.
• The run is stopped.
• Pressure is released.
• High voltage application is stopped.
Agilent 7100 Capillary Electrophoresis System User Manual
163