LAN Configuration

Configuration Switch

Configuration Switch

The configuration switch can be accessed on the right rear side of the
instrument. The 7100 CE is shipped with switches 7 and 8 set to ON. This
means that the instrument is set to a fixed default IP: 192.168.254.11
To perform any LAN configuration, SW1 and SW2 must be set to OFF.
N O T E
Table 17
Initialization ('init') mode
Link configuration
64
Factory default settings
Agilent 7100 Capillary Electrophoresis System User Manual
Using default, switches 7 and 8 set to ON.
Speed and duplex mode determined by
auto-negotiation