Agilent Technologies 7100 User Manual page 254

Capillary electrophoresis system
Hide thumbs Also See for 7100:

Advertisement

10
Appendix
Setting Up a Test Sample Run
Preconditioning
CE Conditioning
• Flush inlet 1, outlet 8 for 5 minutes
• Wait 5 minutes
• Flush inlet 2, outlet 8 for 5 minutes
• Flush inlet 6, outlet 8 for 10 minutes
CE Injection
• Inject by pressure, 50 mbar, 4 seconds
Timetable
• Change voltage 0.2 minutes, 30 kV
Postconditioning
• No postconditioning used
254
Agilent 7100 Capillary Electrophoresis System User Manual