Troubleshooting and Diagnostics

CE Problem Solving

CE Problem Solving
The problem solving section in the instrument online help assists you in
solving the most common failures. Nevertheless, a short list of problems are
featured here.

Power On Fail

The instrument will boot up after turning the instrument on by pressing the
power on switch. A green LED in the power switch is illuminated. If the
instrument power on fails, the green control LED in the power switch may not
light up.
• Check the external power supply, power cords and plugs for proper
• Check the line fuses next to the power plug on the right rear of the
• Check fuses 1 to 6 on the main electronic board. A blown fuse is indicated
162
connection.
instrument.
by a small red LED visible through a hole in the metal shield of the
instrument.