Using Stored
When initialization mode "Using Stored" is selected, the parameters are taken
from the non-volatile memory of the instrument. The TCP/IP connection will
be established using these parameters. The parameters were configured
previously by one of the described methods.
Figure 5
Using Default
When "Using Default" is selected, the factory default parameters are used
instead. These parameters enable a TCP/IP connection to the LAN interface
without further configuration, see
A direct connection of a computer and the 7100 instrument is the entry point
to change the stored parameters of the instrument. The default parameters of
the 7100 instrument are permanently stored and cannot be changed. Using
them, this allows you to establish a LAN connection at any time when the init
mode "Using Default" is configured.
Figure 6
Using the default address in your local area network may result in network problems. Take
N O T E
care and change it to a valid address immediately unless you plan a direct LAN connection
to stand-alone PC.
Agilent 7100 Capillary Electrophoresis System User Manual
Using Stored (principle)
Using Default (principle)
Initialization Mode Selection
Table 19
on page 68.