Table 24
Sleeve Color
Black
Green
Red
Blue*
Yellow
* Use the blue alignment interface for 100 or 150 µm id capillaries
1 Make sure the color of the sleeve in the alignment interface matches the
2 Remove the protective plastic cap that covers the detection window.
The capillary is very fragile at that point and it is important to keep the detection
N O T E
window clean.
• Do not touch the capillary at the detection window or the capillary tips, where the
• Clean the detection window and the capillary tips as described. Put the insertion
Colored sleeve
Alignment interface
Capillary insertion tool
Figure 28
Agilent 7100 Capillary Electrophoresis System User Manual
Alignment interfaces
Part Number
G7100-60150
G7100-60210
G7100-60230
G7100-60310
G7100-60330
color of the plastic stop on the capillary.
polyimide is burnt off.
tool on a flat surface, for example, your lab bench.
Alignment interface and insertion tool
Using the Agilent 7100 Capillary Electrophoresis System
*
Used For
25 µm id capillaries with extended light path
50 µm id straight capillaries
50 µm id capillaries with extended light path
75 µm id straight capillaries
75 µm id capillaries with extended light path