To Compare A Frequency Response Measurement To Synthesized Results - Agilent Technologies 35670A Operator's Manual

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35607A
Operator's Guide
To compare a frequency response measurement to synthesized
results
1
Transfer the curve fit table to a synthesis table. See the previous task.
2
Press [
] [
Disp Format
3
Press [
] [
Active Trace
Press [
] [
Meas Data
FREQUENCY RESPONSE
4
Press [
] [
Active Trace
5
Press [
] [
Analys
SYNTHESIS
6
Press [
] [
Scale
MATCH X SCALE
7
To overlay the results,
press [
] [
Disp Format
The analyzer puts the synthesized frequency response on the active trace. To keep the
original measurement data on Trace A, put the synthesized data on Trace B.
If you overlay the results in step 3, the vertical scaling for trace B is automatically set
to match trace A.
].
UPPER/LOWER
]
A
].
].
B
] [
START SYNTHESIS
] [
TO TRACE A
].
SINGLE FRNT/BACK
].
].
Curve Fit Option 1D3
16-39

Hide quick links: