Synthesized Residual
Measurement using
Beatnote Cal

Procedure

Figure 8-12
This calibration option requires two synthesizers of which one must be
tunable such that a beatnote can be acquired. For the system to calibrate, the
beatnote frequency must be within the following ranges.
Table 8-5
Carrier Frequency
<500 kHz
<5 MHz
<50 MHz
<250 MHz
>250 MHz
or 1/2 the frequency range of the configured analyzer, or
whichever is lower.
1. Connect circuit as per
connections.

Synthesized Residual Measurement using Beatnote Cal

2. Offset the carrier frequency of one synthesizer to produce a beatnote for
cal.
Agilent Technologies E5500 Phase Noise Measurement System 8-19
Residual Measurement Fundamentals
The Calibration Options
Beatnote Frequency Ranges
Beatnote Frequency
Range
10 Hz to 10 kHz
10 Hz to 100 kHz
10 Hz to 1 MHz
10 Hz to 10 MHz
10 Hz to 50 MHz
Figure 8-12 on page
8-19, and tighten all