Operating Instructions
DB-067 E
RadEye G20
RadEye G20-ER
RadEye G20-10
RadEye G20-ER10
X-Ray Gamma Survey Meter

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific RadEye G20-ER

  • Page 1 Operating Instructions DB-067 E RadEye G20 RadEye G20-ER RadEye G20-10 RadEye G20-ER10 X-Ray Gamma Survey Meter...
  • Page 3 REVISIONS SHEET: Rev. Rev. Dept. Name Rev. Cat. Explanation state resp. page 10.12.07 RM&P-E > V 1.32 03.03.08 RM&P-E cpl. 03.04.08 RM&P-E Diagram 9-2 28.07.08 RM&P-EH cpl. > V 1.50 Low Battery warning 12.10.09 RM&SI-E 0-7, back C Name of company 02.02..11 RM&SI-E 4-8, 4-9,...
  • Page 4: Table Of Contents

    CONTENTS: Introduction........1-1 Installation and start-up ....2-1 Scope of delivery ............2-1 Inserting the batteries..........2-2 Mounting of the protection sleeve ......2-4 Switching the unit on (example refers to G20-10, calibrated in rem/h)..2-5 Configuration ........3-1 Menu functions ............3-1 3.1.1 Menu structure...........3-2 3.1.2 Main menu............3-4 3.1.3...
  • Page 5 Scaler mode...............3-9 Alarm thresholds.............3-10 Setting alarm thresholds..........3-10 Operation.......... 4-1 Audible single pulse indication and finder mode..4-1 Alarm indication ............4-2 Additional information ..........4-3 Key Lock ..............4-3 Earphone ..............4-4 Alarm latching ............4-5 Show alarm ...............4-6 Text Info ..............4-7 Display rotation............4-7 4.10 Set Date and Time ..........4-8 4.11 Battery type ............4-8 4.12...
  • Page 6 4.13 Source test ............4-10 4.14 Change menu language ........4-12 Measurement units ......5-1 Functional test ......... 6-1 Functional test............6-1 Failure indication ............6-2 Overload indication...........6-3 Response test with test adapter .........6-4 Low Battery warning ..........6-5 Calibration expiration date........6-6 PC configuration......7-1 Ways to connect to a PC ...........7-1 7.1.1 Wireless connection via Bluetooth ....7-1...
  • Page 7 Starting the program ..........7-5 7.4.1 RadEye G20 Device Parameters......7-5 7.4.2 Measurements............7-6 7.4.3 Creating a Measurement File......7-8 7.4.4 Select serial interface........7-10 Configuration ............7-11 7.5.1 General configuration ........7-12 7.5.2 User defined text..........7-13 7.5.3 Mode setting ............7-14 7.5.4 Source test ............7-15 7.5.5 Alarm setting ...........7-16 7.5.6 Bluetooth ............7-17 7.5.7...
  • Page 8 The RadEye shows an error message in the display’s top line............8-3 The RadEye is not found in the software....8-3 RadEye lost date and time settings ......8-5 RadEye’s battery status is lower than expected ..8-6 History data shows the wrong time and date ....8-6 Maintenance ........
  • Page 9 11.3 Test adapter for RadEye G20 (4254948) ....11-2 11.4 Desktop holder (425067060).......11-3 11.5 IR connection cable serial (4254029)....11-3 11.6 IR connection cable USB (4254026) ....11-4 11.7 Bluetooth battery cover (425067087) ....11-5 11.8 RadEye car- and wall holder with accumulator charging circuit.......11-5 11.8.1 Accessories for dash board mounting of car holder 42506/7065...
  • Page 10 Technical data ......12-1 12.1 RadEye G20-10 and RadEye G20-ER10 ....12-1 12.2 RadEye G20 and RadEye G20-ER......12-8 12.3 RadEye inductive charger: ........12-11 12.4 Firmware revisions..........12-12 V 1.32................12-12 V 1.50................12-12 V 1.51................12-12 V 2.04................12-12 V 2.05................12-12 V 3.01................12-12 Annex ........... 13-1 13.1...
  • Page 11: Weee Compliance

    Diagram 12-1: Energy dependence for ambient dose equivalent H*(10) ............12-5 Diagram 12-2: Relative Gamma response RadEye G20 and RadEye G20-ER ............12-5 Diagram 12-3: Angular response, relative to Cs-137 ....12-6 Diagram 12-4: Angular response, normalized.......12-6 Diagram 12-5: Angular response relative to Cs-137 .....12-7 Diagram 12-6: Angular response, normalized.......12-7...
  • Page 12: Safety Instructions

    SAFETY INSTRUCTIONS The RadEye G20 is suited to perform highly accurate dose measurements. It is however not intended for use as a legal personal dose meter. The RadEye B20 is not well suited for pulsed radiation. There will be a significant under estimation at dose rates during pulse of more than 0.2 R/h.
  • Page 13: Introduction

    SI-standard units of ambient dose equivalent H*(10) (= “deep dose”) and show a flat response from 17 keV to 1.3 MeV. RadEye G20 and RadEye G20-ER are energy compensated according to roentgen units (40 keV to 1.3 MeV). In this context it is important to note, that users may select “R/h, Sv/h or rem/h”...
  • Page 14 A real time clock is provided to add a time stamp to all buffer data. The characteristic feature of the RadEye G20 is the use of sophisticated low power technology components and mi- croprocessor based fully automatic self checks. No mainte- nance is required.
  • Page 15: Installation And Start-Up

    2. Installation and start-up Scope of delivery The RadEye G20 instruments are delivered in a box together with two AAA cells and an operating manual. Ample space is provided for a data cable or a test adapter. Ordering information for accessories see chapter 11 RadEye G20 DB-067 E Iw/Tr/Ff 26.11.2012...
  • Page 16: Inserting The Batteries

    The two AAA-Alkaline cells as delivered with the instrument allow about to 900 h with RadEye G20/G20-10 respectively 700h with RadEye G20-ER/G20-ER10 of normal operation. AAA size rechargeable batteries can be used as well. • Switch off the measuring instrument.
  • Page 17 The measured values in the history memory remain stored. The real time clock for time stamp of history values and log- book continue operation, if batteries are exchanged without delays. If Real Time Clock is set, actual time and date is dis- played for 3 s (see chapter 2.4).
  • Page 18: Mounting Of The Protection Sleeve

    Mounting of the protection sleeve The rubber protection sleeve improves ruggedness to mechani- cal shocks. For mounting of the sleeve first put the instrument into the bot- tom of the sleeve. Then pull lower edges of the sleeve, one after the other into its right position.
  • Page 19: Switching The Unit On (Example Refers To G20-10, Calibrated In Rem/H)

    Switching the unit on (example refers to G20-10, calibrated in rem/h) To switch on the RadEye G20, keep the “On/▼” button pressed for at least one second. The sound generator (beeper) is initi- ated. After switching the unit on, the RadEye G20 starts working with the parameters previously selected (operation mode, cali- bration factor, alarm thresholds etc.).
  • Page 20 With the setting of the history recording interval of 120 s, the last 52 hours of operation will be stored in the history memory. The parameters (except calibration parameters) can be changed with the help of RadEye buttons and the optional PC-program “RadEye.exe”...
  • Page 21: Configuration

    3. Configuration Menu functions To enter the operating menu, press the “Menu” key. Scrolling through the single menu options is effected by the “▲/Info” and “▼/On” keys. The display returns to its initial default setting in case no key has been activated for more than 10 seconds.
  • Page 22: Menu Structure

    The meaning of the Menu key may change with the selected menu. The meaning is shown on the bottom of the display. : Edit values Change : Switching a function on and off : Select a default display mode Select : Confirmation of an action : Exit menu Exit...
  • Page 23 MENU SUBMENU DESCRIPTION OF THE FUNCTION DISPLAY DISPLAY ACTIVATED BY THE “◄/MENU” KEY Clear Dose Clears the accumulated dose. Settings Battery type Selection of battery type: rechargeable or non- rechargeable batteries. Correct low battery warning depends on this setting. Autosend If activated, the instrument sends a data string peri- odically to the infrared port.
  • Page 24: Main Menu

    3.1.2 Main menu The main menu offers the following displays: The above illustration depicts all menu options for the main menu possible for dose rate mode. 3.1.3 Measuring unit This submenu is used to select the measuring unit for ratemeter and scaler measurements.
  • Page 25: Operation Mode

    RadEye G20-ER. However, for convenience reasons, the user may select this unit for RadEye G20-10 and RadEye G20-ER10 with the understanding that the measured quantity is still dose equivalent H*(10) and that the relation to the calibration is set as 1 R = 1 rem = 10 mSv.
  • Page 26: Settings

    3.1.6 Settings In this submenu it is possible to select between audible single pulse indication and finder mode. A “ ” shows the active con- figuration. With “Autosend” the RadEye sends a data string periodically to the infrared port. “Batt.type” is used to set the correct threshold for warning message “LOW BATT”.
  • Page 27: Ratemeter

    Ratemeter The standard display set by factory is indication of the actual dose rate. Measurement Heart beat value Battery indication Bar graph overview Alarm Alarm threshold 1 threshold 2 Pressing the ▲/Info key shows upon the click: the accumulated dose (rem, R or Sv) is shown. It displays the accumulated dose and the remaining time until the dose alarm 1 is reached, assuming the current dose rate will persist.
  • Page 28: Trend Indication

    After 10 seconds or after 3 click standard display with dose rate indication is activated again. click click 3.2.1 Trend indication A trend indication is given, if the measured dose rate is increasing or decreasing. Increasing dose rate Decreasing dose rate For accurate measurements, only readings without trend indica- tion should be used.
  • Page 29: Scaler Mode

    Scaler mode To enter the scaler mode select in the submenu „Scaler“. The scaler mode starts with the parameter set in submenu „Scaler Parameter“. Measure- ment value Preset time Elapsed time Action of ▲/Info key To start the measurement, press the ▲/Info key. While the meas- urement is in progress “Stop”...
  • Page 30: Alarm Thresholds

    Alarm thresholds There are two alarm thresholds each allocated to dose rate and dose. In order to avoid dose alarms while using the instrument exclu- sively as a rate meter, the dose alarms can be set to the maximum level. Configuration of the alarms is possible via infrared interface or via the menu.
  • Page 31 Example: The dose rate alarm threshold must be changed. Press “Menu” key, and ▲/▼keys until is selected. Alarm Dose Rate Then enter change menu by press- ing “Change” key. Edit value by pressing ▲/▼, select digit by ◄/► keys Pre unit “µ” and “m” may be cho- sen as well.
  • Page 33: Operation

    4. Operation Audible single pulse indication and finder mode With the single pulse indication being selected, each pulse of the detector generates a short audible signal emitted by the beeper. An audible alarm signal caused by exceeding the alarm threshold is not given while single pulse indication or finder mode is active.
  • Page 34: Alarm Indication

    Finder mode: In the Finder mode, the sound frequency depends on the pulse rate of the detector. The more detector pulses above the count rate that is present at the moment of the activation of the finder mode are captured by the detector, the higher the tone. The finder mode must be enabled in the menu.
  • Page 35: Additional Information

    The alarm tone and vibrator are acknowledged by a short key depression, the LED remains pulsing. The alarm is extinguished, when the first alarm threshold is remained under. Additional information In addition to the basic display, using the menu options, requests for further information can be started such as e.g.
  • Page 36: Earphone

    It is recommended to lock the keys when wearing the unit in the holster. Thus reduced battery time because of additional power consumption by illumination of the LCD or other unintended operations is avoided. Unlocking is performed according to the LCD instructions upon pressing any key: Press left key first, then lower key and then right key.
  • Page 37: Alarm Latching

    Alarm latching Via PC-program it is possible to configure an alarm latching. Alarm latching means the alarm is stored for the configured time from 0 s (= off) to 9999 s. A latched alarm is indicated by an “M” near by the alarm indica- tor.
  • Page 38: Show Alarm

    Show alarm By selecting the menu point “Show alarm” the alarms stored in the alarm log are displayed together with the time of alarm and the actual time. Actual date and time of the Real Time Clock Stored alarm Date and time of the stored alarm By pressing the ▲...
  • Page 39: Text Info

    Text Info Via PC program “RadEye.exe” it is possible to place text infor- mation in the RadEye. This text information can be displayed with the menu function “Text Info”. Display rotation If enabled, a short press on the “On/▼” key flips the display. If the “On/▼”...
  • Page 40: Set Date And Time

    4.10 Set Date and Time Setting of date and time takes place via menu function Setting → Set Date/Time. The setting uses the format YYYY-MM-DD hh:mm. The fields are selected with the keys ◄/►. With the keys ▲/▼ the Year, month, day, hour or minute can be changed.
  • Page 41: Btcom Cover

    4.12 BTcom cover For detailed installation and operation procedure refer to BTcom cover manual DB-076 E which is delivered together with the Bluetooth battery lid. The BTcom cover applies to the RadEye switched off. With menu item “BTcom”, the user may switch on / off the BTcom cover.
  • Page 42: Source Test

    Start the source test without source. The RadEye check the back- ground value. If the background is lower than 0.15 µSv/h the user is asked to place the source. High range test is only performed with RadEye G20-ER and G20-ER10. 4-10 DB-067 E RadEye G20 Iw/Tr/Ff 26.11.2012...
  • Page 43 Date and time of last success- ful source test Start source test by placing the Low range test source High range test Terminate source test with “►” key After placing the source, source test is started automatically. During measurement the display shows the actual measurement value.
  • Page 44: Change Menu Language

    At the end of source test, is displayed. If every test is successful the actual timestamp is stored and displayed together with the message “Ready”. If source test fails, the message “Inspection” is displayed. Source test is successful Source test fails. RadEye should be sent to inspection 4.14 Change menu language...
  • Page 45: Measurement Units

    5. Measurement units Direction of maximum response To detect hidden radiation sources, the alarm threshold must be set to the lowest value that does not produce fail alarms. This is typical a value of 0.2 to 0.3 µSv/h (20 – 30 µrem/h, µR/h). Due to it’s almost 100 times higher efficiency for gamma emit- ters, the RadEye PRD is more suited for this task, than the RadEye G20.
  • Page 46 The user may as well select R/h as a displayed measuring unit with the understanding that 1 R/h is simply set equal to 1 Rem/h and that consequently energy dependant deviations to the expo- sure rate (and air kerma rate) will occur. The RadEye G20-10 and G20-ER10 is factory calibrated to the ambient dose equivalent rate H*(10) in a depth 10 mm for Cs-137 gamma radiation (662 keV).
  • Page 47: Functional Test

    6. Functional test When the instrument is switched on, it will show zero dose rate until the first detector pulse occurs. The radiation meter performs continuous self-check routines. A complete failure of the detector during operation will be indicated after 1 minute on the LCD and be announced by the beeper. The same applies to the battery voltage.
  • Page 48: Failure Indication

    Failure indication In case of a failure the beeper generates a sharp single pulse every 32 s. The corresponding failure message is displayed in the LCD: Error high voltage generation No detector pulse within 128 seconds EEPROM with calibration data shows EEPROM Read or EEPROM Write error.
  • Page 49: Overload Indication

    Maximum dose rate RadEye G20 200 mR/h RadEye G20-10 200 mrem/h 2 mSv/h RadEye G20-ER 10 R/h RadEye G20-ER10 10 rem/h 100 mSv/h The dose value is marked with an overload indication, if a dose rate overload occurred since the last dose reset.
  • Page 50: Response Test With Test Adapter

    Response test with test adapter The test-adapter 42549/48 offers a convenient way to verify the radiation response performance of the RadEye G20. This “check source” contains 50 g of natural Lu2O3 (which corresponds to approximately 2400 Bq Lu-176). Due to the low resulting gamma dose rate this yields a net dose rate of the RadEye G20-10 of ap- proximately 0.25 µSv/h.
  • Page 51: Low Battery Warning

    Low Battery warning If the battery voltage is below the configured threshold (see chap- ter 4.11) the following warning appears: The beeper generates a single pulse every 32s. This acoustic warning can be suppressed by pressing the alarm acknowledge key. After 8h this warning comes up again. The battery needs to be changed.
  • Page 52: Calibration Expiration Date

    Calibration expiration date If the expiration date is enabled and date is expired, the following warning message is shown at start up of the RadEye: If the expiration date is enabled the RTC is not set (e.g. after change of batteries) the following warning message is shown at start up.
  • Page 53: Pc Configuration

    7. PC configuration Ways to connect to a PC A RadEye unit can be connected to a PC either by either serial TM 1 interface, USB interface or by Bluetooth 7.1.1 Wireless connection via Bluetooth Details about the Bluetooth communication are provided through the manual DB-076 E delivered with the optional Bluetooth battery cover.
  • Page 54: Cable Connection

    7.1.2 Cable connection It is recommended to use the optional RadEye desktop holder to align the RadEye’s IR window with the IR window of a connection cable. The optional serial adapter cable is used to connect to a RS232 port of a PC. This serial port is selected directly in the RadEye.EXE software (see also chapter 7.4.4).
  • Page 55: Connection To A Pc

    Using the USB connection requires the installation of driver software that will generate a virtual serial port that can be selected in the RadEye.EXE software. The driver is delivered on the RadEye.EXE software installation CD ROM. Installation of a connection cable Using the hardware provided and without exerting too much force, attach the connections of the adapter cable to the PC and to the RadEye desktop holder.
  • Page 56 The following steps will show how to install the USB-driver under Windows XP. Power on your computer and boot to Windows. Run the driver setup program “RadEye-USB.exe”. Click Next to continue and start the installation. Wait until the InstallShield Wizard informs you that driver is successfully installed.
  • Page 57: Installation Of The Optional Radeye.exe Software

    Installation of the optional RadEye.EXE soft- ware Open the file “Setup.EXE” on the optional RadEye.EXE Software installation CD Rom and follow the installation guideline. The installation generates a RadEye.EXE Icon on the desktop. Starting the program The icon can be used to start the software RadEye.exe by dou- ble click.
  • Page 58: Measurements

    Figure 7-1: Main Window 7.4.2 Measurements After pressing the button Read Values the current Count rate, dose rate and dose are displayed in this frame. With button Start the measured value is read from the unit at a certain polling interval. Use the scroll bar to define the polling interval.
  • Page 59 The current measured value is added at the right-hand side, and the diagram is shifted to the left. Up to 100 measured val- ues can be represented. The graduation of the ordinate is automatically adjusted to the measured values supplied by the unit.
  • Page 60: Creating A Measurement File

    7.4.3 Creating a Measurement File The actual measured values Count rate and Dose rate that are shown in the "Measurement" display field can be stored in a measurement file. To do this, open the File menu, click on Open Logfile..., and enter path and name.
  • Page 61 With the buttons Start and Stop together with the polling in- terval the replay can be controlled. Close Replay switches back to accessing measurement values via infrared interface. The measurement logfile is an ANSI text file with columns that are separated by <TAB>. This enables this file to be read easily into other programs (such as Excel) where the data can be processed.
  • Page 62: Select Serial Interface

    7.4.4 Select serial interface Via Configuration / Com settings... menu another window is opened from which you may select the corresponding inter- face. Figure 7-3: Select serial port If the connection shall be established with the optional Blue- tooth battery cover, then option “Connected via Bluetooth” must be activated.
  • Page 63: Configuration

    Configuration The configuration of the RadEye can be protected with a password. Figure 7-4: Password Setting No password is set during the first installation of the software. A reminder window is activated, if the configuration window is opened. Figure 7-5: Password reminder RadEye G20 DB-067 E 7-11...
  • Page 64: General Configuration

    On the Configuration / RadEye G20… menu, the following parameters can be modified: • Unit • Time interval of the history • Temperature display • Acoustic indication • Alarm level for all basic displays • Signaling types • Additional surveillances •...
  • Page 65: User Defined Text

    On the “Generals” tab, the user may set the used filter, the time interval for the history memory (1...43200s), the kind of the acoustic rate indication (single pulse, finder 4.1), type of battery, menu language and temperature. 7.5.2 User defined text Figure 7-7: Menu selection On the tab “Text”...
  • Page 66: Mode Setting

    7.5.3 Mode setting Figure 7-8: Mode setting On the “Mode“ tab, the user may set the operation mode, the physical unit, parameters for background measurement and parameters for scaler mode. 7-14 DB-067 E RadEye G20 Iw/Tr/Ff 26.11.2012...
  • Page 67: Source Test

    7.5.4 Source test Figure 7-9: Source test parameter This tab is used to provide the function “source test” with the right data. Nuclide: This name is prompted in function “source test” Reference dose rate: dose rate of this nuclide at reference date Reference Date: reference date for this source Half life: Half life time of this nuclide.
  • Page 68: Alarm Setting

    7.5.5 Alarm setting Figure 7-10: Alarm setting The “Alarm” tab offers the user the possibility of making the alarm settings for the single basic displays. Furthermore, the options on this tab allow beeper and additional alarm monitor- ing processes to be configured. 7-16 DB-067 E RadEye G20...
  • Page 69: Bluetooth

    7.5.6 Bluetooth Figure 7-11: Bluetooth Here the user may configure the BTcom battery cover. The frame “Device” informs about firmware version, serial number and the physical address of the BTcom cover (not the RadEye B-20!). The Parameter “Blue LED on” relates to the blue LED at the backside of the battery cover that shows the connection status.
  • Page 70: Menu Configuration

    7.5.7 Menu configuration Figure 7-12: Menu selection On the “Functions” tab, the functions available to the user can be selected to speed up access to frequently needed functions. For example, if LED alarm is always active and rate level indication is never used, these functions can be hidden for the user.
  • Page 71: History

    History Via the Options / History menu, the values stored in the Rad- Eye G20 data memory can be read out, represented in a x/t diagram and saved to the hard disk of the computer. These data subsequently can be read in and further processed in a spreadsheet program.
  • Page 72 For ratemeter data the blue line shows the mean values, the red line the maximum value within the time interval. If batteries are removed, the time reference is lost. In the dia- gram time information is only provided for history values with valid timing entries.
  • Page 73: Logbook

    Logbook Changes in configuration, occurring alarms and errors are logged in a buffer. These saved events can be read out via Options / Logbook… . The logbook is shown as a table, and can be saved to PC hard disc or printed. The logbook has a maximum of 250 data sets.
  • Page 75: Trouble Shooting

    8. Trouble Shooting RadEye doesn’t power on 8.1.1 Press “On” for at least one second Keep the button “ ” pressed until the LED and the sound is active. If the button is pressed only shortly, then the start up routine is not activated.
  • Page 76: Reading Data From The Start Up Screen

    Reading data from the start up screen The start up screen shows important data like the last calibra- tion, the current voltage and the firmware version. This screen stays active as long as the “ ” button is kept pressed during the power on of the RadEye device. The RadEye doesn’t show the menu items as described in the manual The items available in the RadEye menu can be disabled or...
  • Page 77: The Radeye Shows An Error Message In The Display's Top Line

    The RadEye shows an error message in the display’s top line Please refer also to chapter 6.2. If one of the following failures is indicated, then the RadEye must be repaired by the Thermo Scientific bench repair team: HV-Error Detect Err EEPROM Err Watchdog* Please contact our Customer Service for more details.
  • Page 78 The RadEye.EXE software offers to search for a connection to the RadEye unit on other available serial ports. If there is a faulty connection between the unit and the com- puter, the program shows the error message “Data transfer to instrument fails”...
  • Page 79: Radeye Lost Date And Time Settings

    It is recommended to use the Windows system tools to deter- mine the occupation of the currently active serial ports. Virtual serial ports are labeled accordingly for identification. ´ Device Manager showing the virtual driver for USB „Prolific“ at COM 4 Open next the serial interface step window in the RadEye.EXE software: Configuration / Com settings...
  • Page 80: Radeye's Battery Status Is Lower Than Expected

    RadEye’s battery status is lower than expected Please check if the correct battery type is selected under “set- tings” / “Batt. type” and change the type in case. Rechargeable batteries supply a voltage of 1.2V each, while Alkaline batter- ies provide 1.5V. Rechargeable batteries offer also a more steep drop from ac- ceptable battery voltage to “battery low”...
  • Page 81: Maintenance

    9. Maintenance Recommended maintenance The RadEye is basically maintenance free. 9.1.1 Source test It is recommended to perform a source test routine with the optional Lutetium test adapter on a regular basis. The source test routine is described in chapter 4.13. Level of usage Period between Source tests Intensive, changing climate...
  • Page 82: Remove Batteries For Storage

    9.1.3 Remove batteries for storage It is highly recommended to remove both AAA batteries from the battery compartment, if the RadEye unit is stored for a longer period (> 1 week) to avoid damage through battery leakage. DB-067 E RadEye G20 Iw/Tr/Ff 26.11.2012...
  • Page 83: Spare Parts

    10. Spare Parts 10.1 Available spare parts Item Part Number Front foil RadEye G20 42506703036 Front foil RadEye G20-ER 42506703038 Front foil RadEye G20-10 42506703037 Front foil RadEye G20-ER10 42506703039 Battery cover complete 425067033 Rubber shock protection 42506850013 (rubber boot)
  • Page 85: Optional Accessories

    11. Optional Accessories 11.1 Earphone for RadEye series (425067037) The usage of the earphone is recommended, if the RadEye is switched into FINDER mode and disturbance through noisy envi- ronment shall be compensated. The earphone is plugged into the RadEye’s socket at bottom of the unit.
  • Page 86: Extension Adapter

    11.2 Extension adapter For measurements at spots with high local dose rate, the extension adapter in conjunction with the RadEye adapter can be used. RadEye adapter 42506/7078 Handle 0.35m 42506/7075 Handle 1.2m 42506/7076 Handle 4m 42506/7077 11.3 Test adapter for RadEye G20 (4254948) This test adapter contains 50g Lutetiumoxide.
  • Page 87: Desktop Holder (425067060)

    11.4 Desktop holder (425067060) Alignment between the RadEye’s IR data interface window and the IR interface of a PC connection cable is insured with the us- age of the desktop holder. Please see chapter 7.1.2 for details. 11.5 IR connection cable serial (4254029) The IR plug of the connection cable can be fixed easily by screws at the desktop holder.
  • Page 88: Ir Connection Cable Usb (4254026)

    11.6 IR connection cable USB (4254026) The IR plug of the connection cable can be fixed easily by screws at the desktop holder. The PC connection plug is a 9 pin sub D type for serial COM ports. Please see chapter 7.1.2 for details. Cable length = 1.2 m 11-4 DB-067 E RadEye G20...
  • Page 89: Bluetooth Battery Cover (425067087)

    11.7 Bluetooth battery cover (425067087) The standard battery lid is replaced with the Bluetooth battery cover. Please see chapter 7.1.1 for details. 11.8 RadEye car- and wall holder with accumulator charging circuit Safety instructions Only use NiMH – accumulators, no primary batteries (Alka- line –manganese, zinc –...
  • Page 90 RadEye car holder : 42506/7065 The RadEye car holder 42506/7065 serves as supporting device and power supply unit for the operation of a RadEye instrument in a vehicle. For convenient mounting to the dashboard one of the mechanical adaptors 42506/7061 ... 64 is required. Together with a power supply unit 42506/7066, this support is also suitable for a cost attractive stationary ambient monitoring.
  • Page 91 The RadEye car- and wall holder combines several functions. (1) Mechanical holding device for instruments of the Rad- Eye-family (2) Accumulator charger unit with temperature controlled charging of NiMH accumulators in the device (3) Infrared interface for direct connection to the serial inter- face of computers (4) Alarm contact for connecting alarm indicators up to a power of 24W (at 24V, see RadEye area monitor...
  • Page 92 Then insert the RadEye into the car holder. Make sure that the protection plug is not mechanically interfering with the charging pin of the car holder. 11-8 DB-067 E RadEye G20 Iw/Tr/Ff 26.11.2012...
  • Page 93: Accessories For Dash Board Mounting Of Car Holder 42506/7065

    11.8.1 Accessories for dash board mounting of car holder 42506/7065 (including charging function) holder 42506/7060 (mechanical holder only) Goose neck adaptor kit 42506/7061 Pivot arm adaptor kit 42506/7062 RadEye G20 DB-067 E 11-9 Iw/Tr/Ff 26.11.2012...
  • Page 94 Knuckle joint adaptor kit 42506/7063 Goose neck adaptor kit with suction cup 42506/7064 11-10 DB-067 E RadEye G20 Iw/Tr/Ff 26.11.2012...
  • Page 95: Accessories For Data Transmission Using The Car Adaptor

    11.8.2 Accessories for data transmission using the car adaptor RS232 connecting cable 2m: SM1685 35223 USB 1.1 to RS232 adapter cable: SM1685 35251 USB 2.0 to RS232 adapter cable: SM1685 35255 11.8.3 Accessories for AC – operation of the car adaptor Power supply with ciga- rette lighter socket 42506/7066...
  • Page 96: Radeye Inductive Charger

    11.9 RadEye inductive charger The RadEye holder serves as supporting device and, in combina- tion with the special battery lid 42506/7034, power supply unit for the operation of a RadEye instrument in a vehicle. For con- venient mounting to the dashboard one of the mechanical adap- tors 42506/7061 …...
  • Page 97: Led Indicators

    Safety instructions 1. Only use NiMH – accumulators, no primary batteries (Alka- line – manganese, zinc – carbon ...) 2. Only use two accumulators of the same type (manufacturer, rated capacity ...) 3. Only charge accumulators having the same discharging state 4.
  • Page 99: Technical Data

    12. Technical data 12.1 RadEye G20-10 and RadEye G20-ER10 Radiation type: Gamma, X-ray radiation Measured Ambient Dose Equivalent H*(10) and Rate quantities: Measuring range: Maximum dose rate RadEye 2 mSv/h G20-10 200 mrem/h RadEye 100 mSv/h G20-ER10 10 rem/h overload indication up to 10 Sv/h Overload display: (1000 rem/h) max.
  • Page 100 Response time typ.: 10 s for background to 100μrem/h (to reach 90 %): (1μSv/h) typ.: 5 s for background to 300μrem/h (3μSv/h) typ.: 2 s for background to 1mrem/h (10μSv/h) 17 keV – 1.3 MeV: error less than ±30% Photon Energy 17 keV –...
  • Page 101 Weight: around 300g including AAA cells and pro- tection sleeve Internal memory: The last 1600 measured values are saved and can be read out via PC program. Max- and mean value of dose rate. The time interval is factory preset to 120s by default. Logbook with 250 entries for changes of configuration, occurring alarms and errors.
  • Page 102 Battery service RadEye G20-10: ≈ 900h using two alkaline AAA cells de- life: pending on the operating mode ≈ 600h using 800mAh NiMH accu (type “Ready tot use”) RadEye G20-ER10: ≈ 700h using two alkaline AAA cells depending on the oper- ating mode ≈...
  • Page 103 Photon Energy [keV] Diagram 12-1: Energy dependence for ambient dose equivalent H*(10) 1000 Photon Energy [keV] Diagram 12-2: Relative Gamma response RadEye G20 and RadEye G20-ER Direction of maximum response and reference point RadEye G20 DB-067 E 12-5 Iw/Tr/Ff 26.11.2012...
  • Page 104 -180 26keV -170 -160 -150 -140 65keV -130 -120 205keV -110 -100 662keV 1250keV lower limit upper limit Diagram 12-3: Angular response, relative to Cs-137 -180 -170 -160 26keV -150 -140 -130 65keV -120 205keV -110 -100 662keV 1250keV lower limit upper limit Diagram 12-4: Angular response, normalized 12-6...
  • Page 105 -180 -170 -160 -150 26keV -140 -130 65keV -120 205keV -110 -100 662keV 1250keV lower limit upper limit Diagram 12-5: Angular response relative to Cs-137 -180 -170 -160 26keV -150 -140 65keV -130 -120 205keV -110 -100 662keV 1250keV lower limit upper limit Diagram 12-6: Angular response, normalized RadEye G20...
  • Page 106: Radeye G20-10 And Radeye G20-Er10

    12.2 RadEye G20 and RadEye G20-ER Gamma, X-ray radiation Radiation type: Measured Exposure Dose Rate quantities: Measuring range: Maximum dose rate RadEye G20 200 mR/h RadEye G20-ER 10 R/h overload indication up to 1000 R/h Overload display: max. ± 10 % in the measuring range...
  • Page 107 Working -20°C ... + 50°C temperature: Storage -25°C … + 50°C temperature: Relative humidity: < 93 % at 35°C non condensing 1,8 ... 3.3 V Operating voltage: Degree of protec- IP 32 according to EN 60 529 tion: EMC: Disturbance emission : EN 61000-6-3 Immunity : EN 61000-6-2 Mechanical shock: Drop onto a concrete surface...
  • Page 108 ≈ 4 mA: normal operation without alarm Power consump- tion: signals and LCD illumination RadEye G20-10: ≈ 1 mA RadEye G20-ER10: ≈ 1,2 mA ≈ 25 mA with illuminated LC display ≈ 18 mA LED alarm ≈ 30 mA acoustic alarm ≈...
  • Page 109: Radeye Inductive Charger

    12.3 RadEye inductive charger: Operating voltage 11,5…15V Current consumption 150…200mA With supply from 14V, charging cur- rent 50mA Accumulator charging 40…52 mA current Charge time 20…24h Accumulator RadEye on, 800mAh, discharged Backlight switched off *) Charge time 18…20h Accumulator RadEye switched off 800mAh, discharged Ambient temperature -20…+50°C Operation...
  • Page 110: Firmware Revisions

    12.4 Firmware revisions V 1.32 First edition. V 1.50 - Low Battery warning - Automatic Filter select V 1.51 - Display error in cpm-Scaler-mode eliminated V 2.04 - New menu item “Bluetooth” - Source test V 2.05 - Sub menu “Alarm indication” is maskable V 3.01 - New firmware for new hardware - Change of menu language via menu item „Language“...
  • Page 111: Annex

    13. Annex 13.1 Factory Settings RadEye G20 / G20-ER Item Factory default Language English Measuring unit Type of battery Battery (Alkaline) Mode Ratemeter Acoustic indication Single pulse indication Keyboard Beep when key pressed Display temperature Disabled Display dose Active Display option “up- Disabled side down”...
  • Page 112 Alarm 1, Dose 1000 R Alarm 2, Dose 1000 R Alarm thresholds read Disabled only Alarm indicator, Enabled Sound Alarm indicator, LED Enabled Alarm indicator, Vi- Enabled brator Alarm latching time Active functions All available functions are active Bluetooth: Power off Enabled at battery low Bluetooth: Blue LED...
  • Page 113: Factory Settings Radeye G20-10 / G20-Er10

    13.2 Factory Settings RadEye G20-10 / G20-ER10 Item Factory default Language English Measuring unit Sv/h Type of battery Battery (Alkaline) Mode Ratemeter Acoustic indication Single pulse indication Keyboard Beep when key pressed Display temperature Disabled Display dose Active Display option “up- Disabled side down”...
  • Page 114 Alarm indicator, Enabled Sound Alarm indicator, LED Enabled Alarm indicator, Vi- Enabled brator Alarm latching time Active functions All available functions are active Bluetooth: Power off Enabled at battery low Bluetooth: Blue LED Enabled Bluetooth: Firewall Enabled Bluetooth: Secure Enabled connection Bluetooth PIN 1234...
  • Page 116 Thermo Fisher Scientific Messtechnik GmbH Post address: Frauenauracher Straße 96 P.O. Box 16 60 D-91056 Erlangen, Germany D-91051 Erlangen, Germany +49 9131 998-0 www.thermoscientific.com/rmp +49 9131 998 475 fax info.rmp.erlangen@thermofisher.com...