Stratasys J750 User Guide
2. Lead the cable out of the printer so that you can close the printer, and
3. Close the printer, while checking that the sensor does not move out of
4. Set the UV meter to measure mJ/cm2.
5. Set the range on the meter to 2,000.
6. Connect the cable from the sensor to the UV meter.
7. Turn on the UV meter, and wait until "0 0 0.0" appears on the display.
8. In the wizard screen, confirm that the sensor is positioned correctly, and
Continue with "UV Measurement " on the next page.
DOC-08040 Rev. E
tape it down to make sure that it does not interfere with the moving print
block.
Figure 4-104 Correct UV-sensor placement
position.
click Next.
4 Operating and Maintaining the Stratasys J750 3D Printer
4-69