Stratasys J750 User Guide
9.  Secure the filter set to the bracket. 
10. Wipe the filters, tubes and surrounding area to remove material leaks and 
11. Replace the back of the materials cabinet.
12. In the Maintenance Counters screen, click Reset and then Close (see 
Replacing the
If the printer exhaust is not connected to an external ventilation system, a 
Odor Filter
built-in activated-carbon filter removes odors from printing materials. This 
filter should be replaced regularly (as necessary) to keep your working 
environment pleasant. This is normally done during the preventive-
maintenance service visit.  
Replacing the
The UV lamps used for curing models have a long, but limited, working life. 
UV Lamps
You can test and adjust their effective power with the UV Calibration wizard . 
If you need to replace a UV lamp, follow these instructions: 
1.  Make sure the printer is in offline mode.
Warning
Before continuing, make sure that the safety interlock is not defeated, and that
the UV lamp is not hot.
2.  Start the UV Lamp Replacement wizard from the Options menu. 
DOC-08040 Rev. E
Figure 4-132 Filter set secured to the bracket
spills. 
Figure 4-124 on page 4-80).
Figure 4-133 Offline mode indicator (red)
The online/offline button at the bottom of the printer interface should be 
red. If not, click it to switch the printer to offline mode.
4 Operating and Maintaining the Stratasys J750 3D Printer