Entering Transducer Parameters

To enter the Transducer option, scroll to the Transducer entry on
the Program Menu and press
to Figure 4-2 below. To step through each parameter (or reach a
particular parameter), press the [
page A-3 of Appendix A, Menu Maps
Note: Refer to Chapter 3 for additional information about
clamp-on gas transducers and fixture configurations.
.
Transducer
Type:
Frequency 2.00
Figure 4-2: The Transducer Option Window
1. The first prompt asks you to select whether you are using a
wetted or a clamp-on transducer.
a. Use the [ ] and [ ] keys to scroll between the two types.
b. Press
[ENTER]
IMPORTANT:
Programming Site Data
[ENTER]
Transducer/Pipe
Pipe
Lining
Wetted
Transducer
MHz
Wedge Ang
Wedge Tmp
Wedge SS
Cancel
[SEL]
or
to confirm the choice.
While wetted transducers are listed as a choice,
the PT878GC is designed for use with Shear
wave clamp-on transducers. Consult GE before
selecting wetted transducers.
. The screen appears similar
] key. Refer to Figure A-3 on
Fluid
Clamp-on
Special
Tw
°
50
°C
25
m/s
1219.2
April 2009
Path
µs
14
OK
4-3