Setting Up Pulse/Code Parameters (cont.)
2. The next prompt asks if you want to send a specific number of
pulses. Press
Note: If you do not select this option, pressing the [
returns the meter to the Transmit Code prompt.
3. The next prompt asks how many pulses (from 0 to 16) you
wish to send.
a. Press
[ENTER]
b. Use the [
recommended number.
c. Press
[ENTER]
4. The final prompt, signal inversion, enables you to invert the
transducer signal. Signal inversion may be necessary if your
readings show unusual jitter, or for certain unusual
combinations of pipe and fluid or pipe and lining. Signal
inversion should be off, unless you have received other
instructions. To change its status, use the [ ] and [ ] arrow
keys to move to the appropriate radio button and press
[ENTER]
.
You have completed entering data in the Signal Setup option.
[F2]
Press
confirming the changes.
[F3]
Press
Operate Mode.
Servicing the PT878GC
[ENTER]
to select or deselect the pulse option.
to open the drop-down list.
] or [
] arrow keys to scroll to the
to confirm your selection.
(Cancel) to return to Operate Mode without
(OK) to confirm the new data and return to
April 2009
] key
9-31