Installing DMP-1 Dampening Material with CFG-V
Series Fixtures (cont.)
7. With a marker, trace around the transducer footprint, as shown
in Figure 3-24 below.
Figure 3-24: DMP-1 Material with
8. Remove the fixture and transducers. Then use a utility knife to
cut out the area under the transducer footprint, and peel the cut
material off the pipe, as shown in Figure 3-25 below.
Note: Remove excess dampening material as soon as possible,
before it adheres to the pipe and becomes difficult to
remove.
Figure 3-25: DMP-1 Material with
9. To remove any residual adhesive, clean the cut-out area with a
rag or a piece of sandpaper, or use varnish remover.
10. Reinstall the fixture and transducers on the pipe.
Installing the Dampening Material, Transducers and Fixtures