Entering A Calibration Factor - GE TransPort PT878GC User Manual

Portable gas flowmeter
Table of Contents

Advertisement

Entering Reynolds Correction (cont.)
e. When you have completed entering values, press
to confirm the table and return to the Correction Factors
window.
You have completed entering data for a Reynolds Correction
table.
To confirm the entries, press
Operate Mode.
To leave the window without confirming the entries and return
to Operate mode, press

Entering a Calibration Factor

The Calibration Factor is used as a reference for the flowmeter
electronics to ensure accurate readings. To enter a Calibration
Factor, return to the Reynolds Correction tab at the top of the
Correction Factors option. Press the [ ] arrow key to reach the
Calibration Factor window. The window appears similar to
Figure 4-26 below. To step through each parameter, press the [ ]
key.
Reynolds Correction
Calibration Factor
Meter K-Factor
Data Source
Figure 4-26: The Calibration Factor Window
Programming Site Data
[F3]
[F2]
(Cancel) or the
Correction Factors
Calibration Factor
On
K Factor
Single
1.0000
Edit Table
Cancel
(OK). The meter returns to
[ESC]
Inputs
Off
Table
Edit Table
OK
April 2009
[F3]
(OK)
key.
4-53