Tocotransducer, Ultrasound Transducer, and
Legplate
Revision D
Cleaning: Tocotransducer, Ultrasound Transducer, and Legplate
CAUTIONS
ABRASION—Do not use abrasive cloth, sharp objects, or
abrasive cleaners.
ALCOHOL—Do not use Alcohol in cleaning solutions.
DISCONNECTION—Detach the transducers/cables/legplate
from the monitor.
IMMERSION—Do not immerse transducers/cables/legplate or
hold under running water.
1.
Dampen a cloth or paper towel with one of the following products; then wring
out until only slightly wet:
Sodium Hypochlorite 5.25 % (Bleach) diluted 10:1
Cidex
Sporicidin
Soap and water
2.
Rub soiled area until clean, taking care not to excessively wet the
tocotransducer diaphragm seal or the contact surface of the ultrasound
transducer.
3.
Use a cloth dampened with water on the contact surface of the ultrasound
transducer and around the seal of the tocotransducer. Do not use a sharp object
which might damage the seal of the tocotransducer.
4.
Dry all accessories with a soft, dry cloth.
170 Series Monitor
2003023-001
8-3